器具を改造せずに実現するロボット低侵襲手術の多軸力覚センシング
Multi-Axis Force Sensing in Robotic Minimally Invasive Surgery With No Instrument Modification
da Vinciの手術器具を改造せず、近位シャフトに6軸光学力センサと新設計カニューレを取り付けて先端の横力・モーメント・軸トルクを計測する手法を提案し、モデルベースとニューラルネットによる校正を検証した。
著者: A. H. Hadi-Hosseinabadi, S. E. Salcudean
分類: cs.RO, cs.SY, eess.SY
原文アブストラクト
This paper presents a novel multi-axis force-sensing approach in robotic minimally invasive surgery with no modification to the surgical instrument. Thus, it is adaptable to different surgical instruments. A novel 6-axis optical force sensor, with local signal conditioning and digital electronics, was mounted onto the proximal shaft of a da Vinci EndoWrist instrument. A new cannula design comprising an inner tube and an outer tube was proposed. The inner tube is attached to the cannula interface to the robot base through a compliant leaf spring with adjustable stiffness. It allows bending of the instrument shaft due to the tip forces. The outer tube mechanically filters out the body forces from affecting the instrument bending behavior. A mathematical model of the sensing principle was developed and used for model-based calibration. A data-driven calibration based on a shallow neural network architecture comprising a single 5-nodes hidden layer and a 5x1 output layer is discussed. Extensive testing was conducted to validate that the sensor can successfully measure the lateral forces and moments and the axial torque applied to the instruments distal end within the desired resolution, accuracy, and range requirements.
関連論文
- 偏心配置イメージングプローブを統合したFBG式三軸力センサ触覚/力覚センシング