定常ハンドアイロボット3.0:網膜下注射のための最適化とベンチトップ評価
Steady-Hand Eye Robot 3.0: Optimization and Benchtop Evaluation for Subretinal Injection
網膜下注射のための協調制御ロボットSHER 3.0を開発し、デルタステージの最適化により先端精度30μm未満、繰り返し精度9.3μmを達成した。
著者: Alireza Alamdar, David E. Usevitch, Jiahao Wu, Russell H. Taylor, Peter Gehlbach, Iulian Iordachita
分類: cs.RO
原文アブストラクト
Subretinal injection methods and other procedures for treating retinal conditions and diseases (many considered incurable) have been limited in scope due to limited human motor control. This study demonstrates the next generation, cooperatively controlled Steady-Hand Eye Robot (SHER 3.0), a precise and intuitive-to-use robotic platform achieving clinical standards for targeting accuracy and resolution for subretinal injections. The system design and basic kinematics are reported and a deflection model for the incorporated delta stage and validation experiments are presented. This model optimizes the delta stage parameters, maximizing the global conditioning index and minimizing torsional compliance. Five tests measuring accuracy, repeatability, and deflection show the optimized stage design achieves a tip accuracy of <30 $\mu$m, tip repeatability of 9.3 $\mu$m and 0.02{\deg}, and deflections between 20-350 $\mu$m/N. Future work will use updated control models to refine tip positioning outcomes and will be tested on in vivo animal models.
関連論文
- 片手補助:単一チャネル内視鏡的粘膜下層剥離術における牽引のための連続体ロボット把持器医療ロボティクス
- 脳血管内治療における臨床医とAIの協働:自律性スペクトラムにわたるロボットナビゲーションのCopilot支援モードに関するin-vitroユーザー研究医療ロボティクス
- 低悪性度脾外傷の止血に向けた監督付き自律双腕吸引・焼灼システム医療ロボティクス
- 言語条件付き行動クローニングによる自律内視鏡ガイダンス医療ロボティクス
- MCP-Net:ロボットガイドワイヤ操作のためのリアルタイム冠動脈認識ネットワークとトポロジー解析医療ロボティクス
- 手術フェーズ認識と視覚追従を統合した自律腹腔鏡制御フレームワーク医療ロボティクス