MRI対応ロボットによる同心腱駆動型心臓カテーテルの自動操作
MR-conditional Robotic Actuation of Concentric Tendon-Driven Cardiac Catheters
MRI環境下で動作する圧電モータ駆動のコンパクトなロボットモジュールを開発し、アブレーションカテーテルとシースの自動操作を実現した。閉ループ制御により平均2mm未満の先端位置誤差で経路追従を達成した。
著者: Yifan Wang, Zheng Qiu, Junichi Tokuda, Ehud J. Schmidt, Aravindan Kolandaivelu, Yue Chen
分類: cs.RO
原文アブストラクト
Atrial fibrillation (AF) and ventricular tachycardia (VT) are two of the sustained arrhythmias that significantly affect the quality of life of patients. Treatment of AF and VT often requires radiofrequency ablation of heart tissues using an ablation catheter. Recent progress in ablation therapy leverages magnetic resonance imaging (MRI) for higher contrast visual feedback, and additionally utilizes a guiding sheath with an actively deflectable tip to improve the dexterity of the catheter inside the heart. This paper presents the design and validation of an MR-conditional robotic module for automated actuation of both the ablation catheter and the sheath. The robotic module features a compact design for improved accessibility inside the MR scanner bore and is driven by piezoelectric motors to ensure MR-conditionality. The combined catheter-sheath mechanism is essentially a concentric tendon-driven continuum robot and its kinematics is modeled by the constant curvature model for closed-loop position control. Path following experiments were conducted to validate the actuation module and control scheme, achieving < 2 mm average tip position error.
関連論文
- 片手補助:単一チャネル内視鏡的粘膜下層剥離術における牽引のための連続体ロボット把持器医療ロボティクス
- 脳血管内治療における臨床医とAIの協働:自律性スペクトラムにわたるロボットナビゲーションのCopilot支援モードに関するin-vitroユーザー研究医療ロボティクス
- 低悪性度脾外傷の止血に向けた監督付き自律双腕吸引・焼灼システム医療ロボティクス
- 言語条件付き行動クローニングによる自律内視鏡ガイダンス医療ロボティクス
- MCP-Net:ロボットガイドワイヤ操作のためのリアルタイム冠動脈認識ネットワークとトポロジー解析医療ロボティクス
- 手術フェーズ認識と視覚追従を統合した自律腹腔鏡制御フレームワーク医療ロボティクス