医療用軟体連続ロボットの閉ループ磁気制御による偏向
Closed-Loop Magnetic Control of Medical Soft Continuum Robots for Deflection
単一の回転可能な永久磁石を用いて磁気軟体連続ロボットの閉ループ偏向制御を行い、外乱下での軌道追従性能を検証した。
著者: Zhiwei Wu, Jinhui Zhang
分類: cs.RO
原文アブストラクト
Magnetic soft continuum robots (MSCRs) have emerged as powerful devices in endovascular interventions owing to their hyperelastic fibre matrix and enhanced magnetic manipulability. Effective closed-loop control of tethered magnetic devices contributes to the achievement of autonomous vascular robotic surgery. In this article, we employ a magnetic actuation system equipped with a single rotatable permanent magnet to achieve closed-loop deflection control of the MSCR. To this end, we establish a differential kinematic model of MSCRs exposed to non-uniform magnetic fields. The relationship between the existence and uniqueness of Jacobian and the geometric position between robots is deduced. The control direction induced by Jacobian is demonstrated to be crucial in simulations. Then, the corresponding quasi-static control (QSC) framework integrates a linear extended state observer to estimate model uncertainties. Finally, the effectiveness of the proposed QSC framework is validated through comparative trajectory tracking experiments with the PD controller under external disturbances. Further extensions are made for the Jacobian to path-following control at the distal end position. The proposed control framework prevents the actuator from reaching the joint limit and achieves fast and low error-tracking performance without overshooting.