生体内実験で検証した脳深部超音波焼灼の熱線量モデリング
Deep Brain Ultrasound Ablation Thermal Dose Modeling with in Vivo Experimental Validation
脳腫瘍治療用の針型超音波焼灼において、有限要素法による熱損傷評価を強化したシミュレーションを構築し、MR温度画像で生体内検証した。
著者: Zhanyue Zhao, Benjamin Szewczyk, Matthew Tarasek, Charles Bales, Yang Wang, Ming Liu, Yiwei Jiang, Chitresh Bhushan, Eric Fiveland, Zahabiya Campwala, Rachel Trowbridge, Phillip M. Johansen, Zachary Olmsted, Goutam Ghoshal, Tamas Heffter, Katie Gandomi, Farid Tavakkolmoghaddam, Christopher Nycz, Erin Jeannotte, Shweta Mane, Julia Nalwalk, E. Clif Burdette, Jiang Qian, Desmond Yeo, Julie Pilitsis, Gregory S. Fischer
分類: physics.med-ph, cs.RO
原文アブストラクト
Intracorporeal needle-based therapeutic ultrasound (NBTU) is a minimally invasive option for intervening in malignant brain tumors, commonly used in thermal ablation procedures. This technique is suitable for both primary and metastatic cancers, utilizing a high-frequency alternating electric field (up to 10 MHz) to excite a piezoelectric transducer. The resulting rapid deformation of the transducer produces an acoustic wave that propagates through tissue, leading to localized high-temperature heating at the target tumor site and inducing rapid cell death. To optimize the design of NBTU transducers for thermal dose delivery during treatment, numerical modeling of the acoustic pressure field generated by the deforming piezoelectric transducer is frequently employed. The bioheat transfer process generated by the input pressure field is used to track the thermal propagation of the applicator over time. Magnetic resonance thermal imaging (MRTI) can be used to experimentally validate these models. Validation results using MRTI demonstrated the feasibility of this model, showing a consistent thermal propagation pattern. However, a thermal damage isodose map is more advantageous for evaluating therapeutic efficacy. To achieve a more accurate simulation based on the actual brain tissue environment, a new finite element method (FEM) simulation with enhanced damage evaluation capabilities was conducted. The results showed that the highest temperature and ablated volume differed between experimental and simulation results by 2.1884{\deg}C (3.71%) and 0.0631 cm$^3$ (5.74%), respectively. The lowest Pearson correlation coefficient (PCC) for peak temperature was 0.7117, and the lowest Dice coefficient for the ablated area was 0.7021, indicating a good agreement in accuracy between simulation and experiment.
関連論文
- 片手補助:単一チャネル内視鏡的粘膜下層剥離術における牽引のための連続体ロボット把持器医療ロボティクス
- 脳血管内治療における臨床医とAIの協働:自律性スペクトラムにわたるロボットナビゲーションのCopilot支援モードに関するin-vitroユーザー研究医療ロボティクス
- 低悪性度脾外傷の止血に向けた監督付き自律双腕吸引・焼灼システム医療ロボティクス
- 言語条件付き行動クローニングによる自律内視鏡ガイダンス医療ロボティクス
- MCP-Net:ロボットガイドワイヤ操作のためのリアルタイム冠動脈認識ネットワークとトポロジー解析医療ロボティクス
- 手術フェーズ認識と視覚追従を統合した自律腹腔鏡制御フレームワーク医療ロボティクス