超音波ガイド下自律穿刺介入ロボットシステムの予備評価
Preliminary Evaluation of an Ultrasound-Guided Robotic System for Autonomous Percutaneous Intervention
超音波画像を用いて肝臓領域の特定から病変の位置推定、針の穿刺までを人手なしで行う自律ロボットシステムを構築し、ファントム実験で5.74±2.70mmの精度を確認した。
著者: Pratima Mohan, Aayush Agrawal, Niravkumar A. Patel
分類: cs.RO
原文アブストラクト
Cancer cases have been rising globally, resulting in nearly 10 million deaths in 2023. Biopsy, crucial for diagnosis, is often performed under ultrasound (US) guidance, demanding precise hand coordination and cognitive decision-making. Robot-assisted interventions have shown improved accuracy in lesion targeting by addressing challenges such as noisy 2D images and maintaining consistent probe-to-surface contact. Recent research has focused on fully autonomous robotic US systems to enable standardized diagnostic procedures and reproducible US-guided therapy. This study presents a fully autonomous system for US-guided needle placement capable of performing end-to-end clinical workflow. The system autonomously: 1) identifies the liver region on the patient's abdomen surface, 2) plans and executes the US scanning path using impedance control, 3) localizes lesions from the US images in real-time, and 4) targets the identified lesions, all without human intervention. This study evaluates both position and impedance-controlled systems. Validation on agar phantoms demonstrated a targeting error of 5.74 +- 2.70 mm, highlighting its potential for accurately targeting tumors larger than 5 mm. Achieved results show its potential for a fully autonomous system for US-guided biopsies.
関連論文
- 片手補助:単一チャネル内視鏡的粘膜下層剥離術における牽引のための連続体ロボット把持器医療ロボティクス
- 脳血管内治療における臨床医とAIの協働:自律性スペクトラムにわたるロボットナビゲーションのCopilot支援モードに関するin-vitroユーザー研究医療ロボティクス
- 低悪性度脾外傷の止血に向けた監督付き自律双腕吸引・焼灼システム医療ロボティクス
- 言語条件付き行動クローニングによる自律内視鏡ガイダンス医療ロボティクス
- MCP-Net:ロボットガイドワイヤ操作のためのリアルタイム冠動脈認識ネットワークとトポロジー解析医療ロボティクス
- 手術フェーズ認識と視覚追従を統合した自律腹腔鏡制御フレームワーク医療ロボティクス