センサレス遠隔中心ずれ推定
Sensorless Remote Center of Motion Misalignment Estimation
腹腔鏡手術ロボットにおいて、力センサを使わずに力推定を行い、遠隔中心(RCM)のずれを動的に評価・最適化するフレームワークを提案した。20mm以上のずれで組織損傷の危険があることを示し、ずれが20mm以上の場合に絶対誤差5mm以内でRCMオフセットを推定できることを実験で確認した。
著者: Hao Yang, Lidia Al-Zogbi, Ahmet Yildiz, Nabil Simaan, Jie Ying Wu
分類: cs.RO, eess.IV
原文アブストラクト
Laparoscopic surgery constrains instrument motion around a fixed pivot point at the incision into a patient to minimize tissue trauma. Surgical robots achieve this through either hardware to software-based remote center of motion (RCM) constraints. However, accurate RCM alignment is difficult due to manual trocar placement, patient motion, and tissue deformation. Misalignment between the robot's RCM point and the patient incision site can cause unsafe forces at the incision site. This paper presents a sensorless force estimation-based framework for dynamically assessing and optimizing RCM misalignment in robotic surgery. Our experiments demonstrate that misalignment exceeding 20 mm can generate large enough forces to potentially damage tissue, emphasizing the need for precise RCM positioning. For misalignment $D\geq $ 20 mm, our optimization algorithm estimates the RCM offset with an absolute error within 5 mm. Accurate RCM misalignment estimation is a step toward automated RCM misalignment compensation, enhancing safety and reducing tissue damage in robotic-assisted laparoscopic surgery.