曲面組織の6DロボットOCTスキャニング
6D Robotic OCT Scanning of Curved Tissue Surfaces
ロボット搭載OCTプローブの完全な6次元ハンドアイキャリブレーション用マーカーを提案し、曲面組織の一貫したスキャンを可能にする。
詳しい要約
1. どんなもの?
2. 先行研究と比べてどこがすごい?
3. 技術・手法の肝は?
4. どうやって有効だと検証した?
5. 議論はある?
6. 次に読むべき論文は?
※ AIが要旨から生成した要約です。正確性は原文をご確認ください。
著者: Suresh Guttikonda, Maximilian Neidhardt, Vidas Raudonis, Alexander Schlaefer
分類: cs.CV, cs.RO
原文アブストラクト
Optical coherence tomography (OCT) is a non-invasive volumetric imaging modality with high spatial and temporal resolution. For imaging larger tissue structures, OCT probes need to be moved to scan the respective area. For handheld scanning, stitching of the acquired OCT volumes requires overlap to register the images. For robotic scanning and stitching, a typical approach is to restrict the motion to translations, as this avoids a full hand-eye calibration, which is complicated by the small field of view of most OCT probes. However, stitching by registration or by translational scanning are limited when curved tissue surfaces need to be scanned. We propose a marker for full six-dimensional hand-eye calibration of a robot mounted OCT probe. We show that the calibration results in highly repeatable estimates of the transformation. Moreover, we evaluate robotic scanning of two phantom surfaces to demonstrate that the proposed calibration allows for consistent scanning of large, curved tissue surfaces. As the proposed approach is not relying on image registration, it does not suffer from a potential accumulation of errors along a scan path. We also illustrate the improvement compared to conventional 3D-translational robotic scanning.