I2RISロボットのための拡張一般化Prandtl-Ishlinskiiヒステリシスモデル
An Extended Generalized Prandtl-Ishlinskii Hysteresis Model for I2RIS Robot
網膜手術支援ロボットI2RISの腱駆動に生じる多段階ヒステリシスを高精度に表現するため、切替機構を導入した拡張一般化Prandtl-Ishlinskiiモデルを提案し、従来モデルより精度が向上することを示した。
著者: Yiyao Yue, Mojtaba Esfandiari, Pengyuan Du, Peter Gehlbach, Makoto Jinno, Adnan Munawar, Peter Kazanzides, Iulian Iordachita
分類: cs.RO
原文アブストラクト
Retinal surgery requires extreme precision due to constrained anatomical spaces in the human retina. To assist surgeons achieve this level of accuracy, the Improved Integrated Robotic Intraocular Snake (I2RIS) with dexterous capability has been developed. However, such flexible tendon-driven robots often suffer from hysteresis problems, which significantly challenges precise control and positioning. In particular, we observed multi-stage hysteresis phenomena in the small-scale I2RIS. In this paper, we propose an Extended Generalized Prandtl-Ishlinskii (EGPI) model to increase the fitting accuracy of the hysteresis. The model incorporates a novel switching mechanism that enables it to describe multi-stage hysteresis in the regions of monotonic input. Experimental validation on I2RIS data demonstrates that the EGPI model outperforms the conventional Generalized Prandtl-Ishlinskii (GPI) model in terms of RMSE, NRMSE, and MAE across multiple motor input directions. The EGPI model in our study highlights the potential in modeling multi-stage hysteresis in minimally invasive flexible robots.
関連論文
- 片手補助:単一チャネル内視鏡的粘膜下層剥離術における牽引のための連続体ロボット把持器医療ロボティクス
- 脳血管内治療における臨床医とAIの協働:自律性スペクトラムにわたるロボットナビゲーションのCopilot支援モードに関するin-vitroユーザー研究医療ロボティクス
- 低悪性度脾外傷の止血に向けた監督付き自律双腕吸引・焼灼システム医療ロボティクス
- 言語条件付き行動クローニングによる自律内視鏡ガイダンス医療ロボティクス
- MCP-Net:ロボットガイドワイヤ操作のためのリアルタイム冠動脈認識ネットワークとトポロジー解析医療ロボティクス
- 手術フェーズ認識と視覚追従を統合した自律腹腔鏡制御フレームワーク医療ロボティクス